|
|
|
Previous Section |
Home |
Next Section |
|
|
|||||||||||||
|
|||||||||||||
|
|
|||||||||||||
![]() |
|||||||||||||
|
|
|||||||||||||
|
|
IntroductionThis is the "Main" routine that is modelled after the HP's MAIN program. It is the window that will most likely be open during the tune-up of the instrument. It displays basic operating conditions and detector signals and lets you change several parameters. All displays are updated once every second. |
||||||||||||
|
|
|||||||||||||
|
|
Beam Off/OnClicking this button will turn the beam on or off. Of course, you could also use the "real button" on the 3f control panel, in which case this "software button" will get updated automatically. This is also true for all the other controls in this window. Try it. |
||||||||||||
|
|
|||||||||||||
Display of Detector ReadingsThis part of the window has three different states, which are discussed individually below: |
|||||||||||||
|
|||||||||||||
|
|
|||||||||||||
Sample Offset Voltage |
|||||||||||||
|
|||||||||||||
|
|
|||||||||||||
|
|
Mass DisplayDisplays the current mass and the corresponding B-field value. Actually, only the B-field value has a real, physical meaning. The mass value displayed is only as good as your mass calibration. You can increase and decrease the mass in steps of 1 amu by clicking on the black triangles. Of course, a mass value can also be entered directly. The B-field value cannot be set directly. (Some people prefer "Dalton" as the mass unit, but it always makes me think of the Dalton brothers from the Lucky Luke comics, so I find it less distracting to use "amu", atomic mass units.) |
||||||||||||
|
|
|||||||||||||
|
|
Periodic TableInstead of manually entering a mass number, you can also simply pick an element from a periodic table. This button will take you there. The details of this function are discussed in Section 5 of this manual. |
||||||||||||
|
|
|||||||||||||
|
|
Mass CalibrationIf something is wrong with the mass calibration, this button should directly take you there to fix what needs to be fixed. Details in Section 6 of this manual. |
||||||||||||
|
|
|||||||||||||
|
|
|||||||||||||
Fast AccessSeveral often-used masses are available as buttons for fast access. Just clicking on any of these buttons changes the magnetic field accordingly. The labels refer to the dominant isotope, i.e. clicking on 'Mg' goes to 24Mg. Note: Alternatively it is possible to use the first function keys on the keyboard for this fast access. The keyboard is labeled with the same element names and pressing one of the function keys is equivalent to pressing these virtual buttons with the mouse. However, the keyboard equivalents only work as long as the Main window is active and in the foreground.
|
|||||||||||||
|
|
|||||||||||||
|
|
Integer vs. Exact MassFor all isotopes -with the exception of 12C- the exact mass is not the same as their integer mass. The problem is that the mass calibration (Section 6) is customarily done in such a way that e.g. 1.000 corresponds to hydrogen and 27.000 corresponds to aluminum. Although this is not technically exact, it is convenient and does not have any adverse side effects. If the mass calibration is done in such a way that all isotopes have integer masses, this should be selected here. This way the 'Fast Access' buttons set the magnetic field to integer values and not to the exact masses. |
||||||||||||
|
|
|||||||||||||
|
|
Detection ModeThis array of button does the same as the corresponding buttons on the 3f control panel. Pressing/clicking buttons on either end will change the detection mode. |
||||||||||||
|
|
|||||||||||||
|
|
Raster SizeThis box displays the current raster size of the primary beam. I have noted that sometimes none of the raster size buttons on the 3f control panel is lit when at the same time this computer raster size display says "500 micron". That should scare you! It means that the raster could on without you knowing it. As far as I can tell there seems to be no way to set the raster size with the computer, we can only display the value chosen on the hardware panel. |
||||||||||||
|
|
|||||||||||||
|
|
Manual Magnet ControlThis allows you to disable the magnet control dial on the 3f control panel. It doesn't always seen to work and I am not quite sure why. |
||||||||||||
|
|
|||||||||||||
|
|
Analysis StatusThis display changes to "Analysis Running" when you have started a measurement. During a measurement all the values displayed in this "Main" window are passive (but not necessarily static). The way this hierarchy works is detailed in Section 2 of this manual. |
||||||||||||
|
|
|||||||||||||
|
|
ExitMakes this window go away. |
||||||||||||
|
|
||
|
FJ Stadermann |
|
|